Measurement of the concentration of the pollutants in the work environment
can be made by monitoring in the Employee's Respiratory Zone, or better
still, by fixing Portable Sampling Unit to the employee's clothing. This is
called Personal Sampling. We offer following models for Personal Sampling :
Model PMP-3B, Model RDS-3 & Model LFP-2.
For enabling valid integrated samples to be taken of the air actually inhaled
by the employee, under normal working conditions, the Personal Sampler is
very useful, and is recommended by the National Institute of Occupational
Safety and Health, and Occupational Safety and Health Administration, U.S.A. |
RESPIRABLE DUST SAMPLER Model RDS - 3
|
In work environment it is important to measure the concentration of dust.
However it is not sufficient to know how much is the suspended particulate matter, but what is the respirable component of it is also required
to be known. This is actually the dust, which will go to the lungs and will
cause ill effect on the human health. From research it is found that dust
particles of size more than 10 micron are unable to reach the lungs. The dust
concentration in the size of particles between 0.4 to 10 micron is required to be
measured.
The equipment essentially consists of :-
* A pump with flow rate of 1.6 LPM
* Sampling Head
|
The Pump : It is lightweight with rechargeable batteries lasting for 8 hours when charged fully. The flow rate adjustment is
available to adjust at 1.6 LPM. The rotameter is optional, but it should be available to adjust the flow rate. Once adjusted,
the rotameter can be disconnected.
The sampling head consist of : Filter paper holder & Cyclone. The filter paper holder is of 47 mm dia in which 0.4 micron filter paper can be kept.
The cyclone construction is such that when the air enters, due to the centrifugal force, particles heavier than 10 micron size
are thrown to bottom.
The cyclone & filter paper holder are connected together such a way that air enters the cyclone, heavy particles are thrown below and air with lighter dust particles travels through the filter paper. Particles between 0.4 micron & 10 micron are
collected on the filter paper. The filter paper is weighed before and after the sampling. Keeping flow rate of 1.6 LPM, sampling is done for 8 hrs.
The concentration of respirable dust is reported in micrograms / cubic meter.
Specifications :
Model No. |
HDS-8 |
PMP-3B |
RDS-3 |
LFP-2 |
Voltage |
6V dc |
4.8 V |
4.8 V |
4.8 V |
Battery |
SMF Lead Acid, 4 AH |
Ni-Cd(4x1.2V),
1500 MAH |
Ni-Cd(4x1.2V),
1500 MAH |
Ni-Cd(4x1.2V),
1500 MAH |
Rotameter |
0 - 5 LPM |
optional |
optional |
optional |
Timer |
8 hrs.
adjustable in step
of 1,4 & 8 hrs. |
- |
- |
- |
Pulse Mode Operation |
Normal, 50% & 25% |
- |
- |
- |
Flow Rate |
5 LPM Max. |
3 LPM Max. |
1.6 LPM Max. |
0.5 LPM Max. |
Flow Adjustment |
Needle Valve |
Needle Valve |
Needle Valve |
Needle Valve |
Mat. Of Construction |
Stainless Steel |
Light Aluminium |
Light Aluminium |
Light Aluminium |
Dimension |
240x110x150 mm. |
115x100x55 mm. |
115x100x55 mm. |
115x100x55 mm. |
Weight |
3.1 Kg. |
0.8 Kg. |
0.8 Kg. |
0.8 Kg. |
|
|