Karl Fischer Titrator
 
" We manufacture Karl Fischer Titrator for moisture analysis
both meeting GLP compliance. "
 
KARL FISCHER TITRATOR MA - 101 C

APPLICATIONS
* Cosmetics * Pharmaceuticals * Organic Chemicals
* Inorganic Chemicals * Petrochemicals * Food & Beverages

OVERVIEW
Karl Fischer titration is an accurate, fast and most efficient method for determining the moisture content in samples. When the sample is titrated in the presence of SO2,I2, and an organic base, moisture from the sample gets extracted in the solvent. However, moisture from gases and liquids gets easily dissolved in methanol. Therefore, specially dried methanol is used as a solvent in Karl Fischer titration.

THE SYSTEM

Constant Polarising Current Source
A Dual Platinum electrode is used across which a constant current is generated by a polarizing current source. For starting titration very low volume of dried methanol is needed in the vessel with a capacity of 150 or 250ml.

Automatic addition of Kf Reagent
Kinetically controlled doses of Karl Fischer reagents are added to the solution in a hermetically sealed vessel as per stop criteria selected and the end-point volume is evaluated automatically. The instrument is continuously monitoring drift in the vessel and neutralizes it automatically.

Reagent standardisation as per USP 23/921
The built in reagent standardization procedure by water (H2O ) or Sodium Tartrate (C4H4Na2O62H2O) for KF Titer / concentration is in compliance with USP 23/921.

Inert liquid handling path
The liquid handling path comprises of Teflon tubing, PTFE three way valve, and a gas tight syringe with Teflon piston. It creates a chemically inert system for this highly sensitive analysis.

Magnetic Stirrer for Homogenous Mixing
The magnetic, adjustable speed stirrer is a part of the system in which a sample is stirred vigorously to extract the moisture efficiently.

Automatic drift compensation
Even very small samples can be analyzed, as we have facility of auto drift compensation. It increases accuracy

Result printouts in various formats
The result, including drift, is printed in tabulated form and is displayed on back lighted liquid crystal display (LCD) screen.

SALIENT FEATURES

* Advanced Microcontroller based userfriendly state-of-the-art product design with alphanumeric splash proof soft touch keyboard. User interactive software in a dialogue mode for easy operation with protection against invalid entries. Quit and back keys for better user interaction. 

* Quick interchangeable burette assembly with intelligent recognition for its volume. Burette calibration factor for correction to true end point volume.

* 51 user methods. Out of which Five standard methods which are not editable / deletable.

* Report printouts giving Sample Id., Sample Name, with Date and Time for authentication. Auto incremented Run Number. Factory entered Customer Name & Instrument Sr. No. for GLP compliance. 

* 3 modes of titration : Blank, Titer, and Sample 

* Result units : % w/w, % w/v & ppm

* Selectable report format ( for GLP Compliance)
* Document report   * Parameter report       * Brief report           * Statistics report      
* Reports can be obtained even after power failure 

* Statistics function for mean, standard deviation and linearity of last five titrations 

* Real Time Clock ( RTC) for date & time printout of titration time. Display of pretitration stir time.

* Built - in standardization procedure for Water /Sodium Tartrate ( as per USP) . Moisture in oil as per ASTM method 

* Quick monitoring, and automatic neutralization of drift into vessel to keep ready for next titration

* End point delay up to 99 sec for the samples releasing moisture slowly

* Data down loading facilities to PC. 21 CFR Part II compliance.

* Microcontroller based variable speed, magnetic stirrer with digital indication.

SPECIFICATION

Burette : 20ml Interchangeable
Burette resolution : 1/20,000 or 0.001 ml .
Filling time : less than 20 sec.
Data Memory : Non volatile
End point Detection  : Voltametric
Measuring Range  : 10µg to 500 mg Water
KF Delivery  : Stepper Motor driven Piston Burette
End Point Potential : Adjustable
End Point Time : 1 to 99 sec
Dosing : Kinetically controlled volumes & delay times. 
Programmability  : 51 User Methods.(5 default) > Result units : %w/w , %w/v, ppm , mg/ml
Drift : Manual or Automatic compensation
Reporting : Drift report, Brief report, Parameter report, report in document format last 3 results in Titer mode & 10 in sample mode in Tabular format.
Interface : PC, Printer and Balance
Power  
Dimension  
Weight  

SPECTRALAB Karl Fischer Titrator Model MA 101C as per specifications (Refer above)

Standard Accessories

* KF Electrode : 1 No.     * KF Vessel (150 ml) : 1 No.          * Manual : 1 No.           * Mains Cord : 1 No.

Optional Accessories  ( Please order as per requirement)
Note : Printer or PC to be procurred separately. However, interface for both is built in.

* Due to continuous R & D the specification & Dimensions are subject to Modifications

|        Back to Main Products Range     |    Back    |       Next       |